EVALUATION OF THE DATEX CD-101 AND GODART CAPNOGRAPH MARK II INFRA-RED CARBON DIOXIDE ANALYSERS

Performance of the Datex CD-101 and Godart Capnograph Mark II infra-red carbon dioxide analysers was investigated. The Datex machine uses a solid-state infra-red detector, while the Godart instrument uses a Luft detector. Both instruments were easy to use and the small size and light weight of the D...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:British journal of anaesthesia : BJA 1980-05, Vol.52 (5), p.551-555
Hauptverfasser: BLACKBURN, J.P., WILLIAMS, T.R.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Performance of the Datex CD-101 and Godart Capnograph Mark II infra-red carbon dioxide analysers was investigated. The Datex machine uses a solid-state infra-red detector, while the Godart instrument uses a Luft detector. Both instruments were easy to use and the small size and light weight of the Datex was an added convenience. Linearity, stability, response time and the effect of nitrous oxide were within specification in both machines except when 10% carbon dioxide was used with the Datex analyser. Linearity of the Godart instrument was considerably better than 1% full scale deflection (fsd) quoted in the specification, while the Datex analyser was within 2% fsd (4% fsd with 10% carbon dioxide).
ISSN:0007-0912
1471-6771
DOI:10.1093/bja/52.5.551