Effects of sample-diluent interactions on surface tension and pressures in continuous flow manifolds
When a plasma sample is aspirated into a continuous flow analyser the internal pressure of the manifold increases. The pressure change is the result of insufficient detergent present to bind all the protein in the sample, and is a potential source of inaccuracy and imprecision in the assay. It is re...
Gespeichert in:
Veröffentlicht in: | Clinica chimica acta 1979-11, Vol.99 (1), p.53-57 |
---|---|
1. Verfasser: | |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | When a plasma sample is aspirated into a continuous flow analyser the internal pressure of the manifold increases. The pressure change is the result of insufficient detergent present to bind all the protein in the sample, and is a potential source of inaccuracy and imprecision in the assay. It is recommended that Brij 35 be used at a concentration 10 times above the binding capacity of the albumin in the sample, to minimise changes in pressure between samples. |
---|---|