Complications with retained transvenous pacemaker electrodes

Out of a series of 1,734 pacemaker patients, the clinical course of 46 patients was reviewed in whom a functionless endocardial electrode was retained. Non-infected electrodes (25 patients) were generally well tolerated without complications, except in one patient who experienced fatal catheter embo...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:The American heart journal 1979-11, Vol.98 (5), p.587-594
Hauptverfasser: Rettig, Gerd, Doenecke, Peter, Sen, Semi, Volkmer, Ingo, Bette, Ludwig
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Out of a series of 1,734 pacemaker patients, the clinical course of 46 patients was reviewed in whom a functionless endocardial electrode was retained. Non-infected electrodes (25 patients) were generally well tolerated without complications, except in one patient who experienced fatal catheter embolism into the pulmonary artery. In cases of infected electrodes (21 patients) a mortality rate of 25% was encountered due to septic complications. Catheter migration was fatal in two out of three patients. It is concluded that entrapped electrode catheters should be removed by thoracotomy if persisting infection is present or if catheter migration has occurred.
ISSN:0002-8703
1097-6744
DOI:10.1016/0002-8703(79)90284-9