The effect of pre-incubation and direct procedure for the alanine transaminase assay in two kits compared

The kinetic measurement of serum alanine transaminase was carried out on two enzyme kits (Roche and Calbiochem), done as a direct method and a pre-incubation method. Variable results from commercial control sera and patient sera by these four procedures were obtained. Pre-incubation enzyme values we...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Clinical biochemistry 1979-08, Vol.12 (4), p.135-137
Hauptverfasser: Buttery, J.E., Milner, C.R., Chamberlain, B.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The kinetic measurement of serum alanine transaminase was carried out on two enzyme kits (Roche and Calbiochem), done as a direct method and a pre-incubation method. Variable results from commercial control sera and patient sera by these four procedures were obtained. Pre-incubation enzyme values were generally lower than the direct values (mean difference 10%), but there were marked differences between the results from the two kits. This finding has important significance especially for laboratories participating in external QC Programmes. Frequent precision checks of the method and a local reference range are therefore necessary.
ISSN:0009-9120
1873-2933