PIV measurements of proximal models with different anastomotic angles

The anastomosis glass models are designed and fabricated to investigate the effect of the anastomotic angles (30° to 90° for both forward and backward flow) on the intimal hyperplasia formation. A test rig is designed and built to perform Particle Image Velocimetry (PIV) measurement on the anastomos...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:International communications in heat and mass transfer 2000-05, Vol.27 (4), p.517-526
Hauptverfasser: Chua, L.P., Yu, S.C.M., Tam, W.P.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The anastomosis glass models are designed and fabricated to investigate the effect of the anastomotic angles (30° to 90° for both forward and backward flow) on the intimal hyperplasia formation. A test rig is designed and built to perform Particle Image Velocimetry (PIV) measurement on the anastomosis models with three Reynolds numbers of 100, 169 and 250. The initial results indicated that there are two low velocity regions, one of them is located at the outer wall of the graft near the joint and the other is located at one diameter down from the joint along the inner wall of the graft. It is found that these low velocity regions are dependent on the Reynolds number and anastomotic angles.
ISSN:0735-1933
1879-0178
DOI:10.1016/S0735-1933(00)00134-2