Torsional properties of implant grade titanium

The purpose of this research was to evaluate the torsional strength and ductility of CP titanium in the as received condition, heat treated below the alpha arrow right beta transition temperature, and glass bead blasted. Results were compared to as-received samples of implant quality 316L stainless...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of biomedical materials research 1990-01, Vol.25 (3), p.281-293
Hauptverfasser: Zardiackas, L D, Disegi, J
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The purpose of this research was to evaluate the torsional strength and ductility of CP titanium in the as received condition, heat treated below the alpha arrow right beta transition temperature, and glass bead blasted. Results were compared to as-received samples of implant quality 316L stainless steel. Little or no effeft of any of the treatments was noted. The torsional strength of all titanium samples was comparable to 316L stainless and the torsional ductility of the titanium was significantly greater than 316L stainless (p < .01).
ISSN:0021-9304