The establishment of the NPL infrared relative spectral responsivity scale using cavity pyroelectric detectors

The establishment of a relative spectral responsivity scale requires the availability of a detector with a relative spectral responsivity which can be measured or estimated. NPL has addressed this requirement by designing a new detector with a measurable relative spectral response profile in the 800...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Metrologia 2006-04, Vol.43 (2), p.S115-S119
1. Verfasser: Theocharous, Evangelos
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The establishment of a relative spectral responsivity scale requires the availability of a detector with a relative spectral responsivity which can be measured or estimated. NPL has addressed this requirement by designing a new detector with a measurable relative spectral response profile in the 800 nm to 24 (mu)m wavelength range. The new detector is based on a large area pyroelectric detector enclosed in a removable gold-coated hemispherical reflector. A number of these cavity pyroelectric detectors were fabricated and three have been selected and are currently being used as the primary standards for the establishment of the NPL relative spectral responsivity scale in the infrared region. This paper summarizes the design and performance characteristics of these cavity pyroelectric detectors and demonstrates how they are used to establish the NPL relative spectral responsivity scales in the 0.8 (mu)m to 24 (mu)m wavelength range. [PUBLICATION ABSTRACT]
ISSN:0026-1394
1681-7575
DOI:10.1088/0026-1394/43/2/S23