Magnetomechanical field computations of a clinical magnetic resonance imaging (MRI) scanner
In this paper, an efficient magnetomechanical calculation scheme based on the finite element method is presented. This scheme is used for the precise forecast of the dynamical behavior of a clinical magnetic resonance imaging scanner. The validity of the computer simulations has been verified by mea...
Gespeichert in:
Veröffentlicht in: | Compel 2003-09, Vol.22 (3), p.576-588 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | In this paper, an efficient magnetomechanical calculation scheme based on the finite element method is presented. This scheme is used for the precise forecast of the dynamical behavior of a clinical magnetic resonance imaging scanner. The validity of the computer simulations has been verified by means of appropriate measurements. Application examples include the optimization of the superconducting magnet regarding the eddy currents and vibrations in its cryostat. |
---|---|
ISSN: | 0332-1649 2054-5606 |
DOI: | 10.1108/03321640310475047 |