Evaluation of Emit methods for the determination of the five major antiepileptic drugs on an automated kinetic analyser
We have evaluated the performance of enzyme-multipled immunoassay methods for the five major antiepileptic drugs on an automated system, the Perkin-Elmer Model KA-150 Kinetic Analyzer. The precision in the normal duplicate mode was found to be in the range of 6% to 10% of all five tests over a typic...
Gespeichert in:
Veröffentlicht in: | Clinical biochemistry 1978, Vol.11 (4), p.139-142 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We have evaluated the performance of enzyme-multipled immunoassay methods for the five major antiepileptic drugs on an automated system, the Perkin-Elmer Model KA-150 Kinetic Analyzer. The precision in the normal duplicate mode was found to be in the range of 6% to 10% of all five tests over a typical working day. All EMIT methods were compared to gas-liquid chromatographic procedures and, in addition, the phenytoin and phenobarbital assays were compared to a liquid-chromatographic method. The phenytoin assay was also compared to RIA and to a manual spectroscopic method. In general, most of the comparison studies resulted in acceptable correlation, although one gas chromatographic method did not correlate very well with the phenytoin and phenobarbital immunoassays. |
---|---|
ISSN: | 0009-9120 1873-2933 |
DOI: | 10.1016/S0009-9120(78)90298-9 |