Pacer-Induced Tachycardia Associated with an Atrial Synchronous Ventricular Inhibited (ASVIP) Pulse Generator

A 68‐year‐old white male underwent permanent pacemaker implantation with an atrial synchronous ventricular inhibited pulse generator (Medtronic model 2409) because of syncope and abnormal H‐V interval of 70 ms. Paroxysmal bouts of pacemaker associated tachycardia were subsequently recorded on severa...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Pacing and clinical electrophysiology 1982-03, Vol.5 (2), p.251-259
Hauptverfasser: Serge Barold, S., TOLENTINO, ALFONSO O., JAVIER, ROGER P., BYRD, CHARLES, SAMET, PHILIP
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A 68‐year‐old white male underwent permanent pacemaker implantation with an atrial synchronous ventricular inhibited pulse generator (Medtronic model 2409) because of syncope and abnormal H‐V interval of 70 ms. Paroxysmal bouts of pacemaker associated tachycardia were subsequently recorded on several occasions, initiated and terminated by spontaneous ventricular premature beats. The mechanism for the occurrence of the tachyarrhythmia is discussed in detail and the functional characteristics of the pulse generator are described. Replacement of the unit with a different pacer device prevented further occurrence of the arrhythmia. (PACE, Vol. 5, March‐April, 1982)