Some pitfalls in the quality control of alkaline phosphatase activity
A study of the pH optimum of alkaline phosphatase activity in calibration and control serum preparations reveals a wide variation in pH optima. It is shown that as small an error of ± 0.05 pH units in preparation of the buffered substrate results in an error in the order of ± 5% in the response betw...
Gespeichert in:
Veröffentlicht in: | Clinica chimica acta 1978-07, Vol.87 (1), p.159-163 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A study of the pH optimum of alkaline phosphatase activity in calibration and control serum preparations reveals a wide variation in pH optima.
It is shown that as small an error of ± 0.05 pH units in preparation of the buffered substrate results in an error in the order of ± 5% in the response between the calibrating serum and patient specimens.
It is recommended that a calibration serum should closely parallel the pH optimum of alkaline phosphatase in human sera and that a quality control serum containing alkaline phosphatase of alternative source to that present in the calibration serum be included with patient samples as a sensitive means of monitoring pH changes in the buffered substrate. |
---|---|
ISSN: | 0009-8981 1873-3492 |
DOI: | 10.1016/0009-8981(78)90070-0 |