Calculation of apparent diffusion coefficients (ADCs) in brain using two-point and six-point methods
We evaluated the relative accuracy of calculating apparent diffusion coefficients (ADCs) of intracranial tissues using a two-point versus a six-point regression technique. Echo planar diffusion-weighted MRI was performed at 1.5 T in three standard locations and in pathologic regions in 10 subjects u...
Gespeichert in:
Veröffentlicht in: | Journal of computer assisted tomography 1998-09, Vol.22 (5), p.792-794 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We evaluated the relative accuracy of calculating apparent diffusion coefficients (ADCs) of intracranial tissues using a two-point versus a six-point regression technique. Echo planar diffusion-weighted MRI was performed at 1.5 T in three standard locations and in pathologic regions in 10 subjects using gradient strengths corresponding to b values of 1, 100, 200, 500, 800, and 1,000 s/mm2. Estimation of ADCs was made using two methods: a nonlinear regression model using measurements from the full set of b values (six-point technique) and linear estimation using b values of 1 and 1,000 only (two-point technique). A high correlation between the two methods was noted (R2 = 0.999), and the mean percentage difference was 0.84%. These results suggest there is little error in estimating brain ADCs using a two-point technique. |
---|---|
ISSN: | 0363-8715 1532-3145 |
DOI: | 10.1097/00004728-199809000-00023 |