Improved internal defibrillation success with shocks timed to the morphology electrogram
A previous retrospective study by our group suggested that shocks timed to the upslope of the shocking lead electrogram improved defibrillation efficacy. The goal of this study was to prospectively determine whether defibrillation threshold could be reduced by use of an algorithm that timed shocks t...
Gespeichert in:
Veröffentlicht in: | Circulation (New York, N.Y.) N.Y.), 1998-08, Vol.98 (8), p.808-812 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A previous retrospective study by our group suggested that shocks timed to the upslope of the shocking lead electrogram improved defibrillation efficacy. The goal of this study was to prospectively determine whether defibrillation threshold could be reduced by use of an algorithm that timed shocks to the upslope of coarse ventricular fibrillation (test treatment) compared with shocks delivered asynchronously after 10 seconds of fibrillation (control treatment).
Ten pigs were instrumented with a 3-lead system for internal defibrillation. Initial estimates of the energy required to achieve defibrillation E50 for both treatments were made by an up/down method. Subsequently, additional shocks at V50+/-10% and V50-20% were given for each treatment to obtain data points at higher and lower intensities. Probability-of-success curves were estimated for both treatments by the best-fit method. Energies required were significantly lower for the timed shocks than for the asynchronous shocks (P |
---|---|
ISSN: | 0009-7322 1524-4539 |
DOI: | 10.1161/01.CIR.98.8.808 |