Isotope dilution-liquid chromatography/electrospray ionization-tandem mass spectrometry for the determination of serum thyroxine as a potential reference method

A new candidate reference method is presented for the determination of thyroxine in serum. The method is based on isotope dilution‐liquid chromatography/tandem mass spectrometry using electrospray for ionization. The internal standard used was 13C6‐thyroxine, sample pretreatment consisted of protein...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Rapid communications in mass spectrometry 1998-08, Vol.12 (16), p.1099-1103
Hauptverfasser: De Brabandere, Veronique I., Hou, Pin, Stöckl, Dietmar, Thienpont, Linda M., De Leenheer, André P.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A new candidate reference method is presented for the determination of thyroxine in serum. The method is based on isotope dilution‐liquid chromatography/tandem mass spectrometry using electrospray for ionization. The internal standard used was 13C6‐thyroxine, sample pretreatment consisted of protein precipitation and a two‐step liquid/liquid extraction procedure, HPLC was performed on a C‐18 column with an eluent containing methanol/water/formic acid (60:40:0.1, by volume), and finally thyroxine and its isotopically labeled analogue were measured in the selected reaction monitoring mode for the transitions m/z 777.7 → 731.7 and m/z 783.7 → 737.7, respectively. The detection limit for thyroxine was 6 pg, the within‐run coefficient of variation was 1.1%. The samples were measured in six‐fold: in duplicate on three independent days. The mean overall coefficient of variation of the method was 1.6%. The new method was evaluated by measuring nine control sera previously determined by an existing ID‐GC/MS method. The differences between the results of the two methods ranged from −1.6% to +3.3%, with a mean of +0.2%. © 1998 John Wiley & Sons, Ltd.
ISSN:0951-4198
1097-0231
DOI:10.1002/(SICI)1097-0231(19980831)12:16<1099::AID-RCM290>3.0.CO;2-J