Drain entrapment and titanium to ceramic head deposition: two unique complications following closed reduction of a dislocated total hip arthroplasty
Postoperative dislocation remains one of the most frequent complications following total hip replacement. In this report, a case is presented that illustrates two potential concerns with postoperative dislocation and subsequent closed reduction. The first complication presented in this report is ent...
Gespeichert in:
Veröffentlicht in: | The Journal of arthroplasty 1998-09, Vol.13 (6), p.713-717 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Postoperative dislocation remains one of the most frequent complications following total hip replacement. In this report, a case is presented that illustrates two potential concerns with postoperative dislocation and subsequent closed reduction. The first complication presented in this report is entrapment of a closed drainage system tube in the joint space following closed reduction. The second complication, transfer of metallic debris to a ceramic femoral head from contact with an acetabular shell during closed reduction, was documented by analysis of a femoral head using scanning electron microscopy and energy dispersive x-ray spectrometry. This report emphasizes the need for the surgeon to express caution when relocating a dislocated hip, particularly when a closed drainage system is used postoperatively. |
---|---|
ISSN: | 0883-5403 |
DOI: | 10.1016/S0883-5403(98)80018-1 |