Preclinical evaluation of intra-aortic balloon pumping for pediatric use

Small intra-aortic balloons (0.75 ml to 5.0 ml) mounted on small catheters (4.0F and 5.0F) have undergone in vitro testing with the Utah Heart Driver and in vivo evaluation with 3 different pumping modules. In small animals with reduced cardiac output from acutely produced mitral regurgitation, intr...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Transactions - American Society for Artificial Internal Organs 1981, Vol.27, p.490-494
Hauptverfasser: Veasy, L G, Blaylock, R C, Daitoh, N, Fukumasu, H, Orth, J L
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Small intra-aortic balloons (0.75 ml to 5.0 ml) mounted on small catheters (4.0F and 5.0F) have undergone in vitro testing with the Utah Heart Driver and in vivo evaluation with 3 different pumping modules. In small animals with reduced cardiac output from acutely produced mitral regurgitation, intra-aortic balloon pumping effectively reduced aortic peak systolic pressure, lowered aortic end-diastolic pressure and reduced left atrial pressure. In addition, the Datascope system most effectively augmented diastolic pressure and increased aortic forward and negative flow as compared with 2 other pumping modules. The total system has been adequately and experimentally evaluated and is ready for clinical use in infants and small children.
ISSN:0066-0078