Preclinical evaluation of intra-aortic balloon pumping for pediatric use
Small intra-aortic balloons (0.75 ml to 5.0 ml) mounted on small catheters (4.0F and 5.0F) have undergone in vitro testing with the Utah Heart Driver and in vivo evaluation with 3 different pumping modules. In small animals with reduced cardiac output from acutely produced mitral regurgitation, intr...
Gespeichert in:
Veröffentlicht in: | Transactions - American Society for Artificial Internal Organs 1981, Vol.27, p.490-494 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Small intra-aortic balloons (0.75 ml to 5.0 ml) mounted on small catheters (4.0F and 5.0F) have undergone in vitro testing with the Utah Heart Driver and in vivo evaluation with 3 different pumping modules. In small animals with reduced cardiac output from acutely produced mitral regurgitation, intra-aortic balloon pumping effectively reduced aortic peak systolic pressure, lowered aortic end-diastolic pressure and reduced left atrial pressure. In addition, the Datascope system most effectively augmented diastolic pressure and increased aortic forward and negative flow as compared with 2 other pumping modules. The total system has been adequately and experimentally evaluated and is ready for clinical use in infants and small children. |
---|---|