Neutral carrier based ion-selective electrode for the determination of total calcium in blood serum

A simple potentiometric determination of the total calcium concentration in human blood serum using a liquid membrane electrode is described. The bound calcium is displaced from its complexes through acidification of the serum sample with an acetate buffer to pH 3.5. A novel neutral carrier based Ca...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Analytical chemistry (Washington) 1981-11, Vol.53 (13), p.1970-1974
Hauptverfasser: Anker, Peter, Wieland, Erich, Ammann, Daniel, Dohner, Rene E, Asper, Robert, Simon, Wilhelm
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A simple potentiometric determination of the total calcium concentration in human blood serum using a liquid membrane electrode is described. The bound calcium is displaced from its complexes through acidification of the serum sample with an acetate buffer to pH 3.5. A novel neutral carrier based Ca super(2+)-selective electrode showing a much higher rejection of H super(+) ions in comparison to previously described carrier membranes as well as classical ion-exchanger membranes permits the Ca super(2+) determination at such a low pH value.
ISSN:0003-2700
1520-6882