A new automated method to analyze urinary 8-hydroxydeoxyguanosine by a high-performance liquid chromatography-electrochemical detector system

A new method was developed to analyze urinary 8-hydroxydeoxyguanosine (8-OH-dG) by high-performance liquid chromatography (HPLC) coupled to an electrochemical detector (ECD). This method is unique because (i) urine is first fractionated by anion exchange chromatography (polystyrene-type resin with q...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of radiation research 2003-06, Vol.44 (2), p.185-189
1. Verfasser: Kasai, Hiroshi
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A new method was developed to analyze urinary 8-hydroxydeoxyguanosine (8-OH-dG) by high-performance liquid chromatography (HPLC) coupled to an electrochemical detector (ECD). This method is unique because (i) urine is first fractionated by anion exchange chromatography (polystyrene-type resin with quaternary ammonium group, sulfate form) before analysis by reverse phase chromatography; and (ii) the 8-OH-dG fraction in the first HPLC is precisely and automatically collected based on the added ribonucleoside 8-hydroxyguanosine marker peak, which elutes 4-5 min earlier. Up to 1,000 human urine samples can be continuously analyzed with high accuracy within a few months. This method will be useful for studies in radiotherapy, molecular epidemiology, risk assessment, and health promotion.
ISSN:0449-3060
1349-9157
DOI:10.1269/jrr.44.185