Design of actively shielded main magnets: an improved functional method
An improved functional approach for designing MRI (magnetic resonance imaging) main magnets with active shielding is presented. By nulling one or two external moments as well as a certain series of internal moments of the magnetic field, new designs with improved shielding in combination with or wit...
Gespeichert in:
Veröffentlicht in: | Magma (New York, N.Y.) N.Y.), 2003-07, Vol.16 (2), p.57-67 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | An improved functional approach for designing MRI (magnetic resonance imaging) main magnets with active shielding is presented. By nulling one or two external moments as well as a certain series of internal moments of the magnetic field, new designs with improved shielding in combination with or without shorter magnet lengths are obtained. The improved method can be employed to design short and practical superconducting magnets at any given field strength. The resulting designs yield the desired field homogeneity inside the region of interest without using superconducting shim coils. This approach requires only a modest amount of computing power. One of the design steps, a contour plot of the continuous current solutions, can be utilized to study stretch goals for favorable design parameters. |
---|---|
ISSN: | 0968-5243 1352-8661 1352-8661 |
DOI: | 10.1007/s10334-003-0012-x |