An efficient HPLC method for the analysis of isomeric purity of technetium-99m-exametazime and identity confirmation using LC–MS
99mTc-exametazime ( 99mTc-d,l-HMPAO, 99mTc-d,l-hexamethylpropyleneamine oxime) is a neutral rather unstable complex of short-lived 99mTc ( t 1/2=6 h) with the d,l-isomer (mixture of d, d- and l, l-isomers) of a bis-amine bis-oxime tetraligand. It is widely used for measurement of regional cerebral p...
Gespeichert in:
Veröffentlicht in: | Journal of pharmaceutical and biomedical analysis 2003-08, Vol.32 (4), p.679-685 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | 99mTc-exametazime (
99mTc-d,l-HMPAO,
99mTc-d,l-hexamethylpropyleneamine oxime) is a neutral rather unstable complex of short-lived
99mTc (
t
1/2=6 h) with the d,l-isomer (mixture of
d,
d- and
l,
l-isomers) of a bis-amine bis-oxime tetraligand. It is widely used for measurement of regional cerebral perfusion in nuclear medicine. The meso-isomer (
d,
l-form) should not be present in a preparation as it is not retained in brain and thus does not provide clinically useful information. Meso-HMPAO is removed from the ligand during the synthesis procedure by repeated recrystallization, but can still be present as impurity in d,l-isomer. Due to the lack of a suitable chromatographic method for analysis of the isomeric purity of
99mTc-exametazime preparations, United States Pharmacopoeia 25 (USP 25) prescribes a biological test in rats for quality control purpose. In this study, we developed a suitable high-performance liquid chromatography (HPLC) method which allows to demonstrate the relative amounts of d,l- and meso-isomer in
99mTc-exametazime and so obviates the need for a biodistribution test in animals as part of the quality control. Due to the low concentrations in which
99mTc-d,l-HMPAO is obtained (typically 2–6 ng/ml), confirmation of the identity of
99mTc-d,l-HMPAO in the monograph of the European Pharmacopoeia is now performed only indirectly by TLC and assessment of its retention time on RP-HPLC. To investigate the potential of radio-LC–MS for assessment of the identity of
99mTc-exametazime,
99mTc-d,l-HMPAO and
99mTc-meso-HMPAO prepared using a Tc-rich eluate were analyzed using a radio-LC–MS system equipped with a time-of-flight mass spectrometer with electrospray ionization. The main peak in the radiometric channel coincided with the molecular ion mass of
99mTc-d,l-HMPAO in the mass spectrometer channel and the measured accurate mass differed only by 0.26 ppm from the theoretical mass. The identity of
99mTc-meso-HMPAO was also confirmed. Thus, radio-LC–MS allowed to obtain strong evidence for the structure of
99mTc-d,l-HMPAO and
99mTc-meso-HMPAO at nanomolar concentration. It is concluded that radio-LC–MS can become a sensitive aid in quality control of “no carrier added” radiopharmaceutical preparations. |
---|---|
ISSN: | 0731-7085 1873-264X |
DOI: | 10.1016/S0731-7085(03)00175-4 |