Adjacent segment motion after a simulated lumbar fusion in different sagittal alignments: A biomechanical analysis
An in vitro biomechanical study of adjacent segment motion (at L3-L4 and L5-S1) after a simulated lumbar interbody fusion of L4-L5 in different sagittal alignments was carried out. To test the hypothesis that an L4-L5 fixation in different sagittal alignments causes increased angular motion at the a...
Gespeichert in:
Veröffentlicht in: | Spine (Philadelphia, Pa. 1976) Pa. 1976), 2003-07, Vol.28 (14), p.1560-1566 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | An in vitro biomechanical study of adjacent segment motion (at L3-L4 and L5-S1) after a simulated lumbar interbody fusion of L4-L5 in different sagittal alignments was carried out.
To test the hypothesis that an L4-L5 fixation in different sagittal alignments causes increased angular motion at the adjacent levels (L3-L4 and L5-S1) in comparison with the intact spine.
Clinical experience has suggested that lumbar fusion in a nonanatomic sagittal alignment can increase degeneration of the adjacent levels. It has been hypothesized that this is the result of increased motion at these levels; however, to the authors' knowledge no mechanical studies have demonstrated this.
Eight fresh human cadaveric lumbar spines (L3-S1) were biomechanically tested. Total angular motion at L3-L4 and L5-S1 under flexion-extension load conditions (7-Nm flexion and 7-Nm extension) was measured. Each specimen was tested intact, and then again after each of three different sagittal fixation angles (at L4-L5): (1) in situ (21 degrees lordosis), (2) hyperlordotic (31 degrees lordosis), and (3) hypolordotic (7 degrees lordosis). The simulated anterior/posterior fusion was performed at L4-L5 with pedicle screws posteriorly, vertebral body screws anteriorly, and an interbody dowel.
The averaged values for flexion-extension motion at L3-L4 were as follows: intact specimen 2.0 degrees, in situ fixation 4.0 degrees, hyperlordotic fixation 1.7 degrees, hypolordotic fixation 6.5 degrees. The averaged values for flexion-extension motions at L5-S1 were as follows: intact specimen 2.3 degrees, in situ fixation 2.6 degrees, hyperlordotic fixation 3.6 degrees, hypolordotic fixation 2.9 degrees.
Hypolordotic alignment at L4-L5 caused the greatest amount of flexion-extension motion at L3-L4, and the differences were statistically significant in comparison with intact specimen, in situ fixation, and hyperlordotic fixation. Hyperlordotic alignment at L4-L5 caused the greatest amount of flexion-extension motion at L5-S1, and the difference was statistically significant in comparison with intact specimen but not in situ fixation or hypolordotic fixation. |
---|---|