NO(2) photometer Based on Solid-State Light Sources

A sensitive NO(2) analyzer, that uses a commercial LED emitting near 450 nm, has been developed. The prototype can measure NO(2) concentrations as low as 1 part in 10(9) as defined by the Environmental Protection Agency. The response time of the system is 1 min. In addition, a demonstration of the u...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Applied optics (2004) 1998-08, Vol.37 (24), p.5590-5595
Hauptverfasser: Fetzer, G J, Miao, L, Chilla, J L, Pikal, J M, Menoni, C S
Format: Artikel
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A sensitive NO(2) analyzer, that uses a commercial LED emitting near 450 nm, has been developed. The prototype can measure NO(2) concentrations as low as 1 part in 10(9) as defined by the Environmental Protection Agency. The response time of the system is 1 min. In addition, a demonstration of the use of external-cavity frequency-doubled diode laser radiation to accomplish similar measurements is reported. However, because of low nonlinear optical conversion efficiencies, comparable detection limits were not possible. Nonetheless, frequency-doubled diode-laser wavelengths have the potential to reach farther into the UV than LED technology, allowing the detection and measurement of numerous other gaseous species of interest.
ISSN:1559-128X