New thermooptical measurement method and a comparison with other methods

A simple and sensitive method of measuring the thermally induced index changes arising from absorption of a laser beam in a low-loss material is described. The sample is placed outside the laser cavity, but at the position of minimum radius of curvature of the wavefront, which is a confocal distance...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Applied Optics 1973-01, Vol.12 (1), p.72-79
Hauptverfasser: Hu, C, Whinnery, J R
Format: Artikel
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A simple and sensitive method of measuring the thermally induced index changes arising from absorption of a laser beam in a low-loss material is described. The sample is placed outside the laser cavity, but at the position of minimum radius of curvature of the wavefront, which is a confocal distance behind the waist. It is estimated that the method is sensitive enough to measure absorption coefficients of the order of 5 x 10(_6) cm(_1) and it is shown experimentally to have good accuracy on low-loss materials. A detailed comparison of sensitivity and accuracy estimates is given for the various published thermal-lens methods for measuring low absorption coefficients.Although the method is illustrated for the thermal effect, it is applicable to other nonlinear index changes induced by laser beams.
ISSN:1559-128X
0003-6935
1539-4522
DOI:10.1364/AO.12.000072