Axicon-type test interferometer for cylindrical surfaces: systematic error assessment
The principle and the alignment aberration functions are described for an axicon-type test interferometer for measuring cylindrical mantle surfaces. Additionally, we show that the derived systematic alignment functions fulfill for reasonably small misalignments the requirements for measurements in t...
Gespeichert in:
Veröffentlicht in: | Applied Optics 1997-05, Vol.36 (13), p.2791-2795 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The principle and the alignment aberration functions are described for an axicon-type test interferometer for measuring cylindrical mantle surfaces. Additionally, we show that the derived systematic alignment functions fulfill for reasonably small misalignments the requirements for measurements in the range of approximately 1/100 of a fringe. We verify this with optical path-length calculations, using ray tracing. |
---|---|
ISSN: | 1559-128X 0003-6935 1539-4522 |