Scatter correction in 3-D PET

Modern multiring positron emission tomographs allow the acquisition of 3-D data sets to increase their sensitivity. A substantial part of this data is due to scattered radiation. The authors describe the experimental dependence of point source scatter distributions on energy window setting, source l...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:IEEE transactions on medical imaging 1994-12, Vol.13 (4), p.649-657
Hauptverfasser: Lercher, M.J., Wienhard, K.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext bestellen
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Modern multiring positron emission tomographs allow the acquisition of 3-D data sets to increase their sensitivity. A substantial part of this data is due to scattered radiation. The authors describe the experimental dependence of point source scatter distributions on energy window setting, source location, and scatter volume in geometries relevant for brain studies. The point source scatter distribution was parametrized accurately by a broad, 2-D Gaussian, which included a shift parameter to account for asymmetry of the scatter medium relative to the source. This parametrization was used to formulate two fast scatter correction algorithms suitable for brain scans. In both algorithms, a 2-D subset of the measured projections was transformed into a scatter projection. An image of the 3-D scatter distribution was reconstructed using 2-D algorithms. It was then subtracted from the total (true+scattered) 3-D image. Both algorithms were implemented in different combinations with the additional attenuation correction and were tested on point source and phantom measurements. It was shown that, for the situation typical for brain scans, reconstructed scatter fractions could be reduced to 5% or less.< >
ISSN:0278-0062
1558-254X
DOI:10.1109/42.363103