T wave oversensing and low percentage of biventricular pacing in cardiac resynchronization therapy
The success of cardiac resynchronization therapy (CRT) depends on a high percentage of ventricular pacing. We present the electrocardiography and electrograms of a patient who underwent an implantable cardioverter-defibrillator (ICD) with CRT implantation showing a low percentage of ventricular paci...
Gespeichert in:
Veröffentlicht in: | Cardiology journal 2009, Vol.16 (6), p.580-581 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The success of cardiac resynchronization therapy (CRT) depends on a high percentage of ventricular pacing. We present the electrocardiography and electrograms of a patient who underwent an implantable cardioverter-defibrillator (ICD) with CRT implantation showing a low percentage of ventricular pacing as a result of T wave oversensing of paced QRS. The patient showed no clinical improvement. We suggest finding a good sensitivity of F waves in induced ventricular fibrillation during ICD implantation in order to overcome the harm of possible oversensing. |
---|---|