A magneto-optic probe for magnetic fluctuation measurements

Results from a proof-of-principle experiment are presented that demonstrate it is possible to construct a completely optical, robust, and compact probe capable of spatially resolved measurements of magnetic field fluctuations smaller than 1 G over a frequency range of 1 Hz-8 MHz in a plasma. In cont...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Review of scientific instruments 2009-10, Vol.80 (10), p.103502-103502-4
Hauptverfasser: Przybysz, W. S., Ellis, J., Thakur, S. Chakraborty, Hansen, A., Hardin, R. A., Sears, S., Scime, E. E.
Format: Artikel
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Results from a proof-of-principle experiment are presented that demonstrate it is possible to construct a completely optical, robust, and compact probe capable of spatially resolved measurements of magnetic field fluctuations smaller than 1 G over a frequency range of 1 Hz-8 MHz in a plasma. In contrast to conventional coil probes, the signal strength is independent of fluctuation frequency and the measurement technique is immune to electrostatic pickup. The probe consists of a high Verdet constant crystal, two polarizers, optical fibers, and a photodetector.
ISSN:0034-6748
1089-7623
DOI:10.1063/1.3238509