Two-color subpicosecond optical sampling technique

We have constructed a laser system for performing pump-probe experiments with independently tunable pump and probe pulses. The cross correlation between the two pulses, which determines the resolution of the system, is measured to be 150 fs. This system uses two regeneratively mode-locked Ti:sapphir...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Optics letters 1992-09, Vol.17 (18), p.1286-1288
Hauptverfasser: KAFKA, J. D, PIETERSE, J. W, WATTS, M. L
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We have constructed a laser system for performing pump-probe experiments with independently tunable pump and probe pulses. The cross correlation between the two pulses, which determines the resolution of the system, is measured to be 150 fs. This system uses two regeneratively mode-locked Ti:sapphire lasers with slightly different repetition rates and automatically scans the time delay between the pump and probe lasers 80 times per second.
ISSN:0146-9592
1539-4794