Untreated 24-h intraocular pressures measured with Goldmann applanation tonometry vs nighttime supine pressures with Perkins applanation tonometry

Aims To compare supine nighttime intraocular pressure measurements with Perkins applanation tonometry to 24-h sitting intraocular pressures with Goldmann applanation tonometry. Methods A prospective, untreated, uncontrolled, observational cohort of qualified consecutive ocular hypertensive or primar...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Eye (London) 2010-07, Vol.24 (7), p.1252-1258
Hauptverfasser: Quaranta, L, Konstas, A G P, Rossetti, L, Garcia-Feijoo, J, O'Brien, C, Nasr, M B, Fogagnolo, P, Demos, C M, Stewart, J A, Stewart, W C
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Aims To compare supine nighttime intraocular pressure measurements with Perkins applanation tonometry to 24-h sitting intraocular pressures with Goldmann applanation tonometry. Methods A prospective, untreated, uncontrolled, observational cohort of qualified consecutive ocular hypertensive or primary open-angle glaucoma patients. Patients underwent sitting intraocular pressure measurements over 24-h by Goldmann and patients had their supine nighttime intraocular pressure measurements by Perkins. Results In 100 completed patients, the mean intraocular pressure at 1000, 2200, 0200 and 0600 hours while sitting was 22.5±3.7 mm Hg, and in the supine position, 23.5±4.3 mm Hg ( P
ISSN:0950-222X
1476-5454
DOI:10.1038/eye.2009.300