Evaluation of the analytical performance of the Nova StatSensor creatinine meter and reagent strip technology for whole blood testing

To evaluate the analytical performance of the Nova StatSensor creatinine meter. Imprecision, linearity, patient correlation and interference studies were completed. Total imprecision was 4.5–9.1%CV and a linear measurement range of 93–863 µmol/L (1.05–9.76 mg/dL) was verified. Whole blood creatinine...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Clinical biochemistry 2010-08, Vol.43 (12), p.1026-1029
Hauptverfasser: Schnabl, Kareena L., Bagherpoor, Sousan, Diker, Pauline, Cursio, Christine, DuBois, Jeffrey, Yip, Paul M.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:To evaluate the analytical performance of the Nova StatSensor creatinine meter. Imprecision, linearity, patient correlation and interference studies were completed. Total imprecision was 4.5–9.1%CV and a linear measurement range of 93–863 µmol/L (1.05–9.76 mg/dL) was verified. Whole blood creatinine results correlated well with laboratory plasma measurements ( R 2 = 0.9328) but exhibited a negative proportional bias. In vitro, high levels of creatine and urea falsely elevated creatinine measurement. The creatinine meter provides reliable measurement across a clinically relevant range and has potential use in point-of-care testing.