Evaluation of the analytical performance of the Nova StatSensor creatinine meter and reagent strip technology for whole blood testing
To evaluate the analytical performance of the Nova StatSensor creatinine meter. Imprecision, linearity, patient correlation and interference studies were completed. Total imprecision was 4.5–9.1%CV and a linear measurement range of 93–863 µmol/L (1.05–9.76 mg/dL) was verified. Whole blood creatinine...
Gespeichert in:
Veröffentlicht in: | Clinical biochemistry 2010-08, Vol.43 (12), p.1026-1029 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | To evaluate the analytical performance of the Nova StatSensor creatinine meter.
Imprecision, linearity, patient correlation and interference studies were completed.
Total imprecision was 4.5–9.1%CV and a linear measurement range of 93–863
µmol/L (1.05–9.76
mg/dL) was verified. Whole blood creatinine results correlated well with laboratory plasma measurements (
R
2
=
0.9328) but exhibited a negative proportional bias. In vitro, high levels of creatine and urea falsely elevated creatinine measurement.
The creatinine meter provides reliable measurement across a clinically relevant range and has potential use in point-of-care testing. |
---|---|