In vivo wear rate of the Birmingham Hip Resurfacing arthroplasty. A review of 10 retrieved components

The wear of 8 femoral and 2 acetabular components of the Birmingham Hip Resurfacing (Midland Medical Technologies Ltd, Birmingham, UK) was measured using a coordinate measuring machine (CMM5; SIP, Geneva, Switzerland). After a median in situ time of 13 months, measurements of the Birmingham Hip Resu...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:The Journal of arthroplasty 2009-09, Vol.24 (6), p.951-956
Hauptverfasser: Witzleb, Wolf-Christoph, Hanisch, Uwe, Ziegler, Joerg, Guenther, Klaus-Peter
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The wear of 8 femoral and 2 acetabular components of the Birmingham Hip Resurfacing (Midland Medical Technologies Ltd, Birmingham, UK) was measured using a coordinate measuring machine (CMM5; SIP, Geneva, Switzerland). After a median in situ time of 13 months, measurements of the Birmingham Hip Resurfacing femoral heads revealed a median volumetric wear of 2.9 mm(3). In one case with a cup abduction angle of 70 degrees , a significantly higher wear of 17.8 mm(3) was found. The 2 acetabular components showed volumetric wear of 5.4 and 27.6 mm(3) after 14 and 15 months in vivo. The wear rates of the investigated femoral components were somewhat, but not significantly, higher than the previously investigated 28-mm Metasul heads (median, 0.8 mm(3); Zimmer GmbH, Winterthur, Switzerland).