Early Failure of a Nonmodular Titanium Femoral Stem After Primary Hip Arthroplasty

Abstract Fully porous-coated cementless implants have a low failure rate and excellent survivorship in primary and revision surgery. Failure secondary to implant fracture is extremely rare with modern implant designs. We report an early fracture of a fully porous-coated femoral stem after primary to...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:The Journal of arthroplasty 2010-02, Vol.25 (2), p.333.e1-333.e5
Hauptverfasser: Azzam, Khalid A., MD, Austin, Matthew S., MD, Sharkey, Peter F., MD
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Abstract Fully porous-coated cementless implants have a low failure rate and excellent survivorship in primary and revision surgery. Failure secondary to implant fracture is extremely rare with modern implant designs. We report an early fracture of a fully porous-coated femoral stem after primary total hip arthroplasty. The implant failed because of a filling mismatch between the proximal and distal femur.
ISSN:0883-5403
1532-8406
DOI:10.1016/j.arth.2008.11.104