Complications Encountered With the Use of Constrained Acetabular Prostheses in Total Hip Arthroplasty
Abstract At our department, 46 constrained acetabular components in 38 patients were placed successively for a period of 4 years. Indications included recurrent dislocation, septic and aseptic loosening with extensive bone loss, tumor surgery with extensive bone resection, and instability due to neu...
Gespeichert in:
Veröffentlicht in: | The Journal of arthroplasty 2010-02, Vol.25 (2), p.287-294 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Abstract At our department, 46 constrained acetabular components in 38 patients were placed successively for a period of 4 years. Indications included recurrent dislocation, septic and aseptic loosening with extensive bone loss, tumor surgery with extensive bone resection, and instability due to neurologic impairment. Because 2 cup failures and 10 dislocations were observed with the constrained devices at 4 to 7 years of follow-up, the authors started to use large-diameter metal-on-metal bearings for similar indications. A series of 36 such bearings in 38 patients revealed only one cup failure and one dislocation at 2 to 4 years of follow-up. Although the 2 series are different and therefore difficult to compare, the authors recommend judicious use of constrained devices because of the high failure rate (26%) and consideration of alternative options such as the use of large-diameter metal-on-metal bearings. |
---|---|
ISSN: | 0883-5403 1532-8406 |