A chemiluminescence sensor for the determination of hydrogen peroxide
A chemiluminescence one-shot sensor for hydrogen peroxide is described. It is prepared by immobilization of cobalt chloride and sodium lauryl sulphate in hydroxyethyl cellulose matrix cast on a microscope cover glass. Luminol, sodium phosphate and the sample are mixed before use and applied on the m...
Gespeichert in:
Veröffentlicht in: | Talanta (Oxford) 2007-06, Vol.72 (4), p.1378-1385 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A chemiluminescence one-shot sensor for hydrogen peroxide is described. It is prepared by immobilization of cobalt chloride and sodium lauryl sulphate in hydroxyethyl cellulose matrix cast on a microscope cover glass. Luminol, sodium phosphate and the sample are mixed before use and applied on the membrane by a micropipette. The calibration graph is linear in the range 20–1600
μg/L, and the detection limit of the method (3
σ) is 9
μg/L. A relative standard deviation of 4.5% was obtained for 100
μg/L H
2O
2 (
n
=
11). The sensor has been applied successfully to the determination of hydrogen peroxide in rainwater. |
---|---|
ISSN: | 0039-9140 1873-3573 |