Prospective randomized comparison of 65%/65% versus 42%/42% tilt biphasic waveform on defibrillation thresholds in humans

The waveform tilt of biphasic shocks yielding the lowest defibrillation threshold (DFT) is not well defined. Some evidence indicates that tilts less than 65% may improve DFTs. In 57 patients undergoing ICD implantation, DFTs were determined with truncated exponential biphasic waveform tilts at 65%/6...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of interventional cardiac electrophysiology 2003-06, Vol.8 (3), p.221-225
Hauptverfasser: SHEPARD, Richard K, DEGROOT, Paul J, PACIFICO, Antonio, WOOD, Mark A, ELLENBOGEN, Kenneth A
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The waveform tilt of biphasic shocks yielding the lowest defibrillation threshold (DFT) is not well defined. Some evidence indicates that tilts less than 65% may improve DFTs. In 57 patients undergoing ICD implantation, DFTs were determined with truncated exponential biphasic waveform tilts at 65%/65% and at 42%/42%. An external defibrillator with custom software was used for testing. The effective capacitance of the defibrillator was 132-microF for both waveforms. DFTs were determined using a binary search method starting with 12 Joules (J). Patients were randomly assigned to initial testing with either one of the two tilts. Thirty patients (Group 1) were tested with a two electrode (active can to RV coil, or SVC coil to RV coil) and 27 patients (Group 2) were tested with a three electrode system (subcutaneous patch or active can + SVC coil to RV coil). Groups 1 and 2 did not differ in age, ejection fraction or antiarrhythmic medications. Group 1 delivered energy DFTs were 10.1 +/- 5.5 J with the 65%/65% tilt and 10.1 +/- 5.9 J for the 42%/42% tilt (p = 0.92). In group 2 the average DFT for the 65%/65% tilt was 8.4 +/- 5.7 J and for the 42%/42% tilt was 8.1 +/- 5.3 J (p = 0.70). There were no significant differences in DFTs for either group. The system impedance for Group 1 was 64 +/- 12 ohms and for Group 2 was 39 +/- 6 ohms (p < 0.0001). We found no differences in DFTs between 65%/65% tilt and 42%/42% tilt using either 2- or 3-electrode defibrillation systems. Further research is needed to optimize waveforms in order to minimize DFTs, which will result in smaller ICDs and/or greater safety margins for defibrillation.
ISSN:1383-875X
1572-8595
DOI:10.1023/A:1023925423580