Linear polyamines as carriers in thiocyanate-selective membrane electrodes
The polyamines, octyl-[2-(2-octylamino-ethylamino)-ethyl]-amine (L 1) and octyl-{2-[2-(2-octylamino-ethylamino)-ethylamino]-ethyl}-amine (L 2), have been used as anion ionophores in PVC-based membrane ion-selective electrodes. Different electrodes were prepared containing L 1, or L 2, and o-nitrophe...
Gespeichert in:
Veröffentlicht in: | Talanta (Oxford) 2006-02, Vol.68 (4), p.1182-1189 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The polyamines, octyl-[2-(2-octylamino-ethylamino)-ethyl]-amine (L
1) and octyl-{2-[2-(2-octylamino-ethylamino)-ethylamino]-ethyl}-amine (L
2), have been used as anion ionophores in PVC-based membrane ion-selective electrodes. Different electrodes were prepared containing L
1, or L
2, and
o-nitrophenyl octyl ether (NPOE) or bis(2-ethylhexyl)sebacate (DOS) as plasticizers. The response of the electrodes was tested in two different buffers, HEPES-KOH (pH 7) and MES-KOH (pH 5.6). Electrodes containing L
1 and L
2 with NPOE (E1 and E2, respectively) showed a Nernstian response for thiocyanate with a good response time. The detection limit, linear range and slope for electrode E1 were 3.8
×
10
−6
mol
dm
−3, 1
×
10
−5 to 1
×
10
−1
mol
dm
−3 and −57.2
mV
decade
−1 at pH 5.6 and 4.47
×
10
−6
mol
dm
−3, 1.95
×
10
−5 to 1
×
10
−1
mol
dm
−3 and −58.1
mV
decade
−1 at pH 7.0. For electrode E2 the detection limit, linear range and slope found were 2.63
×
10
−6
mol
dm
−3, 7.94
×
10
−6 to 1
×
10
−1
mol
dm
−3 and −58.5
mV
decade
−1 at pH 5.6 and 1.23
×
10
−5
mol
dm
−3, 7.95
×
10
−5 to 1
×
10
−1
mol
dm
−3 and −46.0
mV
decade
−1 at pH 7. In contrast, electrodes containing DOS as plasticizers gave only response at pH 5.6 (detection limit, linear range and slope at pH 5.6 were 3.16
×
10
−5
mol
dm
−3, 1
×
10
−4 to 1
×
10
−1
mol
dm
−3 and −52.6
mV
decade
−1). Selectivity coefficients for different anions with respect to thiocyanate were calculated. The electrode E2 at pH 5.6 was also used for the determination of SCN
− by potentiometric titrations with Ag
+ ions with good results. The electrode E2 was also used to determine concentrations of thiocyanate in biological samples. |
---|---|
ISSN: | 0039-9140 1873-3573 |
DOI: | 10.1016/j.talanta.2005.07.022 |