A flow cell optosensor for lead based on immobilized gallocynin in chitosan membrane

Gallocynin immobilized in chitosan membrane has been studied as a sensor element of an optical sensor for lead using a flowing system. By using this set up, lead in solution has been determined in the concentration range from 1.0×10 −1 to 1.0×10 3 ppm with a detection limit of 0.075 ppm. The standar...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Talanta (Oxford) 2002-09, Vol.58 (3), p.459-466
Hauptverfasser: Yusof, Nor Azah, Ahmad, Musa
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Gallocynin immobilized in chitosan membrane has been studied as a sensor element of an optical sensor for lead using a flowing system. By using this set up, lead in solution has been determined in the concentration range from 1.0×10 −1 to 1.0×10 3 ppm with a detection limit of 0.075 ppm. The standard deviation of the method for the repeatability of lead detection at a concentration of 100 ppm was found to be 2.10%. The response of the sensor was reproducible and can be regenerated by using acidified saturated KNO 3 solution. Interference from foreign ions was also studied at 1:1 mole ratio of Pb(II):foreign ions.
ISSN:0039-9140
1873-3573