Utilization of standards generated in the process of thermal decomposition chemically modified silica gel for a single point calibration of a GC/FID system
The utilization of a multicomponent gaseous standard mixture, containing CO and CO 2 and obtained by thermal decomposition of a so-called immobilized compound, for a single point calibration of a GC/FID system are described. The generation of such a mixture takes place as a result of thermal decompo...
Gespeichert in:
Veröffentlicht in: | Talanta (Oxford) 1997-09, Vol.44 (9), p.1551-1561 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The utilization of a multicomponent gaseous standard mixture, containing CO and CO
2 and obtained by thermal decomposition of a so-called immobilized compound, for a single point calibration of a GC/FID system are described. The generation of such a mixture takes place as a result of thermal decomposition of a sample of chemically modified silica gel placed in a heated chamber of thermal desorber coupled with the device being calibrated via a catalytic methanizer. The mean amount of the analyte librated from unit mass of the gel was 0.71 mg g
−1 (RSD = 3%) for carbon monoxide and 0.86 mg g
−1 (RSD = 3%) for carbon dioxide. |
---|---|