Simultaneous determination of uric acid and creatinine in urine by an eco-friendly solvent-free high performance liquid chromatographic method
A simple, rapid, and eco-friendly analytical method for simultaneous determination of uric acid and creatinine in urine applying high performance liquid chromatography (HPLC) is described. After dilution, de-protein, and filtration, the sample solution was injected to separate the species with C-18...
Gespeichert in:
Veröffentlicht in: | Talanta (Oxford) 2002-10, Vol.58 (4), p.711-717 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A simple, rapid, and eco-friendly analytical method for simultaneous determination of uric acid and creatinine in urine applying high performance liquid chromatography (HPLC) is described. After dilution, de-protein, and filtration, the sample solution was injected to separate the species with C-18 column by an eluent containing 0.05 M ammonium phosphate buffer at pH 7.4. An UV detector was used to monitor the separation of species at 235 nm. Optimum conditions for separation and detection were investigated. Results indicated that under optimized condition measurements were achieved within 13 min. The detection limits were 0.127 and 0.156 μg ml−1 for uric acid and creatinine respectively. The recovery was 95% (0.57% RSD) for uric acid and 99.2% (0.98% RSD) for creatinine, from five measurements. Real urine specimens were tested. |
---|---|
ISSN: | 0039-9140 1873-3573 |
DOI: | 10.1016/S0039-9140(02)00377-6 |