The use of internal standards in ICP-MS

Careful study of the matrix effect in ICP-MS showed that, in all cases studied, the magnitude of the signal suppression or enhancement depends in a regular way on the mass number. Hence, accurate correction for non-spectral interferences is only possible using an internal standard with mass number c...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Talanta (Oxford) 1992-07, Vol.39 (7), p.737-742
Hauptverfasser: Vanhaecke, F., Vanhoe, H., Dams, R., Vandecasteele, C.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Careful study of the matrix effect in ICP-MS showed that, in all cases studied, the magnitude of the signal suppression or enhancement depends in a regular way on the mass number. Hence, accurate correction for non-spectral interferences is only possible using an internal standard with mass number close to that of the analyte element(s). It is also shown that using an internal standard with mass number close to that of the analyte improves the precision. For both cases, the ionization energy of the internal standard seems to be of no or only secondary importance. To obtain optimal precision and accuracy, the internal standard should be selected as close in mass number as possible to that of the analyte element(s). When a number of elements over a considerable mass range are to be determined, several internal standards have to be used.