Application of electron probe micro-analysis to the estimation of chlorine in alumina-based heterogeneous catalysts

Chlorine in alumina-based catalysts has been determined with a scanning electron microscope attached to an energy dispersive x-ray analyser (SEM-EDX). The method is less time consuming compared to conventional methods involving sample dissolution followed by titrimetry, absorption spectrophotometry...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Talanta (Oxford) 1992, Vol.39 (1), p.17-19
Hauptverfasser: Koshy, V.J., Rao, K.V., Kalpana, G., Garg, V.N.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Chlorine in alumina-based catalysts has been determined with a scanning electron microscope attached to an energy dispersive x-ray analyser (SEM-EDX). The method is less time consuming compared to conventional methods involving sample dissolution followed by titrimetry, absorption spectrophotometry or ion chromatography. The spectrometer is calibrated with laboratory prepared standards. This technique is found suitable for the estimation of chlorine in the range 0.1–1.0% (w/w) with a relative standard deviation < 10% for chlorine levels above 0.2%
ISSN:0039-9140
1873-3573
DOI:10.1016/0039-9140(92)80044-E