Prospective evaluation of a new self-expanding plastic stent for inoperable esophageal strictures
High stent cost is considered the major drawback of self-expanding metal stents for dysphagia palliation in patients with inoperable esophageal strictures. We report our experience with a self-expanding plastic (Polyflex) stent, the cost of which is half that of the metal stents. Between September 1...
Gespeichert in:
Veröffentlicht in: | Surgical endoscopy 2003-06, Vol.17 (6), p.891-895 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | High stent cost is considered the major drawback of self-expanding metal stents for dysphagia palliation in patients with inoperable esophageal strictures. We report our experience with a self-expanding plastic (Polyflex) stent, the cost of which is half that of the metal stents.
Between September 1999 and April 2001, 16 dysphagic patients (15 men; mean age, 69.4 +/- 14.5 years; range, 49-100 years; mean dysphagia score, 3.31 +/- 0.6) with esophageal strictures who underwent Polyflex stent placement (insertion device diameter, 12-14 mm; postexpansion inner stent diameter, 16-21 mm; stent length, 9, 12, and 15 cm) were studied prospectively. The strictures were caused by postsurgical recurrence of gastric/esophageal cancer at the anastomotic site in five patients, primary esophageal cancer in four patients, esophagocardia junction cancer in four patients, metastatic mediastinal lymph nodes from a primary lung cancer invading the esophagus in 1 patient, and benign peptic stricture in two elderly patients. All the patients were prospectively followed until death.
Stent insertion was technically successful in 12 patients (75%). Stent placement failed in four patients (25%) because of failure to pass the delivery catheter across the stricture in three patients and failure of the stent to open in one patient. Early and late stent migration occurred in two patients and 1 patient, respectively. Tumor overgrowth occurred in 1 patient. The mean dysphagia score 7 days after stent placement was 1.1 +/- 0.9. Mean survival was 100.6 +/- 71.2 days (range, 8-225 days).
Self-expanding Polyflex stents are safe and effective for inoperable esophageal strictures and have an acceptable technical success rate. Further experience, better selection criteria, and design improvements should improve results. |
---|---|
ISSN: | 0930-2794 1432-2218 |