In vitro validation of an ultrasonic flowmeter in order to measure the functional residual capacity in newborns
Ultrasonic transit-time airflow meters (UFM) allow simultaneous measurements of volume flow V'(t) and molar mass MM(t) of the breathing gas in the mainstream. Consequently, by using a suitable tracer gas the functional residual capacity (FRC) of the lungs can be measured by a gas wash-in/wash-o...
Gespeichert in:
Veröffentlicht in: | Physiological measurement 2003-05, Vol.24 (2), p.355-365 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Ultrasonic transit-time airflow meters (UFM) allow simultaneous measurements of volume flow V'(t) and molar mass MM(t) of the breathing gas in the mainstream. Consequently, by using a suitable tracer gas the functional residual capacity (FRC) of the lungs can be measured by a gas wash-in/wash-out technique. The aim of this study was to investigate the in vitro accuracy of a multiple-breath wash-in/wash-out technique for FRC measurements using 4% sulphur hexafluoride (SF6) in air. V'(t) and MM(t) were measured with a Spiroson SCIENTIFIC flowmeter (ECO Medics, CH) with 1.3 ml dead space. Linearity of airflow and MM were tested using different tidal volumes (V(T)) and breathing gases with different O2 and SF6 concentrations. To determine the accuracy of FRC measurements SF6 wash-in and wash-out curves from four mechanical lung models (FRC of 22, 53, 102 and 153 ml) were evaluated by the Spiroson. For each model five measurements were performed with a physiological V(T)/FRC ratio of 0.3 and constant respiratory rate of 30 min(-1). The error of measured V(T) (range 4-60 ml) was |
---|---|
ISSN: | 0967-3334 1361-6579 |
DOI: | 10.1088/0967-3334/24/2/311 |