Artifacts caused by cochlear implants with non-removable magnets in 3T MRI: phantom and cadaveric studies

The aim of this study was to evaluate artifacts produced by cochlear implants (CI) during 3.0 Tesla (T) magnetic resonance imaging of the brain using different sequences on phantom and cadaveric specimens. A phantom and three cadaveric specimens with CIs were imaged using a 3.0 T clinical scanner. A...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:European archives of oto-rhino-laryngology 2009-12, Vol.266 (12), p.1885-1890
Hauptverfasser: Majdani, Omid, Rau, Thomas S., Götz, Friedrich, Zimmerling, Martin, Lenarz, Minoo, Lenarz, Thomas, Labadie, Robert, Leinung, Martin
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The aim of this study was to evaluate artifacts produced by cochlear implants (CI) during 3.0 Tesla (T) magnetic resonance imaging of the brain using different sequences on phantom and cadaveric specimens. A phantom and three cadaveric specimens with CIs were imaged using a 3.0 T clinical scanner. Artifacts were analyzed quantitatively and according to the sequence used. Different brain regions were evaluated for image distortion and limitation of diagnostic significance. In cadaver studies, all sequences generated signal-void areas around the implant. In T2-weighted sequences, additional periodic shadowing was discovered. Anatomical structures of the brain on the contralateral side of the CI were for the most part undistorted. At 3T, artifacts around CIs with non-removable magnets compromise image quality of the nearby brain regions and diagnosis of brain lesions is limited. In the contralateral hemisphere, diagnostic accuracy is only marginally limited.
ISSN:0937-4477
1434-4726
DOI:10.1007/s00405-009-0994-8