Valve-in-Valve Implantation Using a Novel Supravalvular Transcatheter Aortic Valve: Proof of Concept
Background Transcatheter valve implantation within degenerated bioprostheses is a potentially promising treatment for high-risk surgical patients. Clinical experience is limited; however, we have shown in vitro that currently available transcatheter aortic valve sizes did not provide acceptable hemo...
Gespeichert in:
Veröffentlicht in: | The Annals of thoracic surgery 2009-12, Vol.88 (6), p.1864-1869 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Background Transcatheter valve implantation within degenerated bioprostheses is a potentially promising treatment for high-risk surgical patients. Clinical experience is limited; however, we have shown in vitro that currently available transcatheter aortic valve sizes did not provide acceptable hemodynamics in small bioprostheses. The objective of this study was to develop a new transcatheter valve that would provide good hemodynamics within degenerated bioprostheses. Methods Supravalvular transcatheter valves were created using a Dacron covered stainless steel stent at the base and trileaflet pericardial leaflets in an open stent above the bioprosthesis. The transcatheter valves were implanted within 19-, 21-, and 23-mm Carpentier-Edwards Perimount bioprostheses with simulated degeneration using BioGlue to achieve a mean pressure gradient of 50 mm Hg. Hemodynamics of valve-in-valve implantation were studied in a pulse duplicator. Results Supravalvular transcatheter valves successfully relieved bioprosthetic stenosis. Acceptable hemodynamics were achieved with a significant reduction in mean pressure gradient of 54.0 ± 3.5 to 9.2 ± 6.3 mm Hg in 23-mm bioprostheses ( p < 0.001), from 49.3 ± 3.1 to 14.4 ± 4.7 mmHg ( p < 0.001) in 21 mm, and from 53.9 ± 3.8 to 28.3 ± 9.8 mm Hg ( p = 0.013) in 19-mm bioprostheses. Effective orifice area after valve-in-valve implantation increased significantly and was comparable to rereplacement with the same size bioprosthesis. Conclusions Valve-in-valve implantation was performed using a novel supravalvular transcatheter valve, which successfully relieved bioprosthetic stenosis. The hemodynamics were comparable with standard surgical valve replacement. Further studies are required to assess device safety and efficacy in patients. |
---|---|
ISSN: | 0003-4975 1552-6259 |
DOI: | 10.1016/j.athoracsur.2009.08.004 |