Optimization of a triple etalon interferometer
The High Resolution Doppler Image (HRDI) is a triple etalon Fabry-Perot interferometer designed to measure Doppler shifts of rotational lines in the O2 atmospheric system from the Upper Atmosphere Research Satellite. These shifts are used to determine wind vectors in the stratosphere and mesosphere....
Gespeichert in:
Veröffentlicht in: | Applied Optics 1987-07, Vol.26 (14), p.2817-2827 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The High Resolution Doppler Image (HRDI) is a triple etalon Fabry-Perot interferometer designed to measure Doppler shifts of rotational lines in the O2 atmospheric system from the Upper Atmosphere Research Satellite. These shifts are used to determine wind vectors in the stratosphere and mesosphere. This paper presents the techniques used to determine the gap thicknesses and reflectivities of the three etalons of the HRDI instrument. The spacings are found to be 1.000, 0.186, and 0.025 cm. These spacings are independent of the reflectivity of the etalons. The reflectivities of the three etalons should be nearly equal to minimize the errors in the wind measurement caused by mistuning of the etalons. The choice of the reflectivity does not strongly influence the statistical error in the wind error when the values are less than about 0.90. |
---|---|
ISSN: | 0003-6935 1559-128X 1539-4522 |
DOI: | 10.1364/AO.26.002817 |