Multiplexed continuous-wave diode-laser cavity ringdown measurements of multiple species

Rapid cavity ringdown measurements of multiple broadband absorbing species (methanol and isopropanol) in gas mixtures have been recorded with two multiplexed continuous-wave distributed-feedback diode lasers operating near 1.4 mum. A measurement sensitivity of 2.4 x 10(-9) cm(-1) for a 4.3-s averagi...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Applied Optics 2000-04, Vol.39 (12), p.2009-2016
Hauptverfasser: Totschnig, G, Baer, D S, Wang, J, Winter, F, Hofbauer, H, Hanson, R K
Format: Artikel
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Rapid cavity ringdown measurements of multiple broadband absorbing species (methanol and isopropanol) in gas mixtures have been recorded with two multiplexed continuous-wave distributed-feedback diode lasers operating near 1.4 mum. A measurement sensitivity of 2.4 x 10(-9) cm(-1) for a 4.3-s averaging time was achieved in a 39.5-cm-long static cell with 99.94% reflectivity mirrors. This corresponds to a water-vapor detection limit of less than 2 ppb (parts in 10(9)) for the strong H(2)O lines near 1.4 mum. The shot-to-shot noise of the decay time constant tau was approximately 0.3-0.7%, and ringdown acquisition rates as great as 900 Hz were achieved.
ISSN:1559-128X
0003-6935
1539-4522
DOI:10.1364/AO.39.002009