Magnetic resonance imaging of pacemakers and implantable cardioverter-defibrillators without specific absorption rate restrictions

The purpose of the current study is to evaluate the safety profile of patients with pacemakers or implantable cardioverter-defibrillators (ICDs) undergoing a medically necessary magnetic resonance imaging (MRI) scan without limitation on peak specific absorption rate (SAR). Recent series suggest tha...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Europace (London, England) England), 2010-07, Vol.12 (7), p.947-951
Hauptverfasser: Mollerus, Michael, Albin, Glenn, Lipinski, Margaret, Lucca, Jill
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The purpose of the current study is to evaluate the safety profile of patients with pacemakers or implantable cardioverter-defibrillators (ICDs) undergoing a medically necessary magnetic resonance imaging (MRI) scan without limitation on peak specific absorption rate (SAR). Recent series suggest that MRI scanning can be performed safely in select patients with pacemakers or ICDs. These studies, though, limited peak SAR. One-hundred and three patients with a total of 240 leads underwent a total of 127 scans of any body landmark using usual protocols with standard peak SAR settings for the scan. No patient was pacemaker dependent. Thresholds were obtained immediately before and after the scan. For all scans, the median (25th and 75th percentiles) peak SAR was 2.5 (1.3, 3.2) W/kg whereas the median scan time was 1650 (1236, 2099) s. Pre- and post-scan pacing thresholds were unchanged [0.7 (0.5, 0.8) vs. 0.6 (0.5, 0.8) V at 0.5 ms, P=NS], though the sensed amplitudes [6.7 (2.9, 11.5) vs. 6.1 (2.9, 11.2) mV, P
ISSN:1099-5129
1532-2092