Simultaneous dual-band ultra-high resolution optical coherence tomography
Ultra-high resolution optical coherence tomography (OCT) imaging is demonstrated simultaneously at 840 nm and 1230 nm central wavelength using an off-the-shelf turn-key supercontinuum light source. Spectral filtering of the light source emission results in a double peak spectrum with average powers...
Gespeichert in:
Veröffentlicht in: | Optics express 2007-08, Vol.15 (17), p.10832-10841 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Ultra-high resolution optical coherence tomography (OCT) imaging is demonstrated simultaneously at 840 nm and 1230 nm central wavelength using an off-the-shelf turn-key supercontinuum light source. Spectral filtering of the light source emission results in a double peak spectrum with average powers exceeding 100 mW and bandwidths exceeding 200 nm for each wavelength band. A free-space OCT setup optimized to support both wavelengths in parallel is introduced. OCT imaging of biological tissue ex vivo and in vivo is demonstrated with axial resolutions measured to be < 2 mum and < 4 mum at 840 nm and 1230 nm, respectively. This measuring scheme is used to extract spectroscopic features with outstanding spatial resolution enabling enhanced image contrast. |
---|---|
ISSN: | 1094-4087 1094-4087 |
DOI: | 10.1364/OE.15.010832 |