In vivo monitoring of blood oxygenation in large veins with a triple-wavelength optoacoustic system
A noninvasive optoacoustic technique could be a clinically useful alternative to existing, invasive methods for cerebral oxygenation monitoring. Recently we proposed to use an optoacoustic technique for monitoring cerebral blood oxygenation by probing large cerebral and neck veins including the supe...
Gespeichert in:
Veröffentlicht in: | Optics express 2007-11, Vol.15 (24), p.16261-16269 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A noninvasive optoacoustic technique could be a clinically useful alternative to existing, invasive methods for cerebral oxygenation monitoring. Recently we proposed to use an optoacoustic technique for monitoring cerebral blood oxygenation by probing large cerebral and neck veins including the superior sagittal sinus and the internal jugular vein. In these studies we used a multi-wavelength optoacoustic system with a nanosecond optical parametric oscillator as a light source and a custom-made optoacoustic probe for the measurement of the optoacoustic signals in vivo from the area of the sheep neck overlying the external jugular vein, which is similar in diameter and depth to the human internal jugular vein. Optoacoustic signals induced in venous blood were measured with high resolution despite the presence of a thick layer of tissues (up to 10 mm) between the external jugular vein and the optoacoustic probe. Three wavelengths were chosen to provide accurate and stable measurements of blood oxygenation: signals at 700 nm and 1064 nm demonstrated high correlation with actual oxygenation measured invasively with CO-Oximeter ("gold standard"), while the signal at 800 nm (isosbestic point) was independent of blood oxygenation and was used for calibration. |
---|---|
ISSN: | 1094-4087 1094-4087 |
DOI: | 10.1364/OE.15.016261 |